• Medical Vacuum Suction System with Buffer Chamber
  • Medical Vacuum Suction System with Buffer Chamber
  • Medical Vacuum Suction System with Buffer Chamber
  • Medical Vacuum Suction System with Buffer Chamber
  • Medical Vacuum Suction System with Buffer Chamber

Medical Vacuum Suction System with Buffer Chamber

Oil or Not: Oil Free
Structure: Rotary Vacuum Pump
Exhauster Method: Entrapment Vacuum Pump
Vacuum Degree: High Vacuum
Work Function: Mainsuction Pump
Working Conditions: No Special Request
Samples:
US$ 2750/Set 1 Set(Min.Order)
| Request Sample
Customization:

Contact Supplier

Diamond Member Since 2020

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Model NO.
CAVS-60
Flow Rate
4-1300m3/H (Per Pump)
Vacuum Range
0.5-500mbar (Absolute Pressure)
Suction Volume
25-630m3/H
Motor Power
0.75-20kw
Terminal Negative Pressure Value
Adjustable Within The Range of 0.02MPa-0.07MPa
Terminal Suction Rate
at Least 30L/Min
System Leakage Rate Per Hour
Max 1%
Transport Package
Wooden Case
Specification
CAVS-60
Trademark
Cape-Golden
Origin
China
HS Code
8414100090
Production Capacity
100

Product Description

Medical Vacuum Suction System with Buffer Chamber

·System Introduction
Medical Vacuum Suction System is based on the vacuum pump unit of the center suction station as the vacuum pressure source. Through the suction of the vacuum pump, the suction pipe reaches the required negative pressure value, and generates suction in the operating room, emergency room, treatment room and the terminal of each ward. It is mainly used to remove sputum, blood, pus and other pollutants from the patient's body.
Medical Vacuum Suction System assembled by Europe vacuum pump and medical gas bacteria filter, with stable vacuum degree, low noise and no pollution. Medical Vacuum Suction System could be started and stopped by manually and automatically, fault alarm timely and comprehensively, and realized signal remote transmission (optional).
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber

·System Components

1. Central Suction Station
It is an independent operation room, composed by vacuum pump unit and vacuum container. Generally including vacuum pump, vacuum tank, negative pressure automatic control cabinet, alarm, sewage tank, negative pressure pipeline and terminal.
Vacuum pump is installed two sets, one operation, and the other standby, to ensure the uninterrupted work.
Vacuum tank is a container to store negative pressure, to prevent the vacuum pump from starting frequently.
The negative pressure automatic control cabinet could start and stop the two units respectively by manual control and automatic control, the two modes switch automatically. When the pressure reaches the alarm pressure value, sound-light alarm is carried out. The low pressure alarm is -0.073mpa, the high pressure alarm is -0.019mpa. It will come into operation after power failure.
The drain tank is used to store the dirt in the suction pipe system, discharge regularly.
The alarm is installed in the duty room, and the sound-light alarm can be heard and seen within the range of 1.5m under the environment of 60dBa noise.
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber

2. Suction Pipe
The piping system from the central suction station to each ward, floor and ward, mainly including vacuum pipe and some pipe elements.

3. Terminals
It is the end of suction line, plug type self-sealing quick joint, connecting the device such as suction device and suction bottle.
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber

·Technical Parameters
Flow rate 4-1300m3/h (per pump)
Vacuum range 0.5-500mbar (absolute pressure)
Suction Volume 25-630m3/h
Motor Power 0.75-20KW
Terminal Negative Pressure Value adjustable within the range of 0.02mpa-0.07mpa
Terminal Suction Rate ≥30L/min
System Leakage Rate per Hour ≤1%

·System Features
High vacuum, stable performance and long service life
No access to water
Compact structure, economic space occupation
Simple operation, easy maintenance, less noise
Multiple units can be realized by PLC joint control
GDMS system can be connected
Comply with European CE EN737-3 standard
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber
Medical Vacuum Suction System with Buffer Chamber


 

Send your message to this supplier

*From:
*To:
avatar Mrs. Yelling
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Contact Supplier

Diamond Member Since 2020

Suppliers with verified business licenses

Manufacturer/Factory
Registered Capital
5000000 RMB
Plant Area
101~500 square meters