Still deciding? Get samples of $ !
Request Sample
|
| Customization: | Available |
|---|---|
| Measure Range: | 10ppm~21%, 100ppm~21%, 1000ppm~21% O2 |
| Sample Gas Flow: | 150~300ml/MIM |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
|
Technical Parameters Table of Oxygen Analyzer
|
|
|
Parameter Category
|
Specific Specifications
|
|
Measurement Range
|
0~10/100/1000ppm, 0~25% (automatic range switching available)
|
|
Accuracy
|
- 0.01~9.99ppm O: ±5% FS
|
|
- 10.0~99.9ppm O: ±3% FS
|
|
|
- 100~1000ppm O: ±2% FS
|
|
|
Repeatability
|
- 0.01~9.99ppm O: ±2.5% FS
|
|
- 10.0~99.9ppm O: ±1.5% FS
|
|
|
- 100~1000ppm O: ±1% FS
|
|
|
Stability
|
- 0.01~9.99ppm O: ±2.5% FS/7d
|
|
- 10.0~99.9ppm O: ±1.5% FS/7d
|
|
|
- 100~1000ppm O: ±1% FS/7d
|
|
|
Response Time
|
T < 60s (at 25°C)
|
|
Power Supply
|
85~264VAC, 50/60Hz
|
|
Signal Output
|
4-20mA (default configuration), 0-20mA, 0-1V, 0-10V; free mode setting available
|
|
Switch Output
|
Two-way alarm; Relay contact capacity: 24V DC, 0.2A
|
|
Communication Interface
|
RS485 (standard)/RS232 (optional)
|
|
Ambient Temperature
|
0~45°C
|
|
Ambient Humidity
|
<80%RH
|
|
Sample Gas Pressure
|
Atmospheric pressure ±10% (gas outlet must be under atmospheric pressure)
|
|
Sensor Type
|
Fuel cell oxygen sensor
|
|
Analyzer Dimension (W×H×D)
|
275mm×190mm×300mm
|
|
Mounting Hole Dimension (W×H)
|
224mm×180mm
|
|
Analyzer Weight
|
Approximately 6.8kg
|
Order knowledge
Measure range
Tested gas pressure
Gas composition and impurity