• Ci-PC36 Dew Point Analyzer
  • Ci-PC36 Dew Point Analyzer
  • Ci-PC36 Dew Point Analyzer
  • Ci-PC36 Dew Point Analyzer
  • Ci-PC36 Dew Point Analyzer
  • Ci-PC36 Dew Point Analyzer

Ci-PC36 Dew Point Analyzer

Place of Origin: Shanghai China (Mainland)
Sensor Life: 20000 Hours
Accuracy: ± 2%Fs
Repeatability: ±1%Fs
Name: Infrared Gas Analyzer
Application: Dew Point Analyzer
Samples:
US$ 1500/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Diamond Member Since 2020

Suppliers with verified business licenses

Beijing, China
Importers and Exporters
The supplier has import and export rights
High Repeat Buyers Choice
More than 50% of buyers repeatedly choose the supplier
Experienced Team
The supplier has 6 foreign trading staff(s) and 3 staff(s) with over 6 years of overseas trading experience
Management Certification
The supplier has quality management system certification, including:
ISO9001:2015 certificate
Medical Device Quality Management System Certification
to see all verified strength labels (16)
  • Overview
  • Product Description
  • Features
  • Product Parameters
  • Applications
  • FAQ
Overview

Basic Info.

Model NO.
CI-PC36
Transport Package
as Per Customer Request
Specification
SGS
Trademark
Cape Golden
Origin
Beijing, China.
HS Code
9027100090
Production Capacity
200 Units Per Month

Product Description

Product Description

CI-PC36 Dew point analyzer

CI-PC36 dew point analyzer is an intelligent dew point analyzer with microprocessor as its core and ultra-thin alumina capacitive sensor as its measuring unit. It has the characteristics of high precision, good stability, comprehensive function and long service life.


Ci-PC36 Dew Point Analyzer

Features
  • With microprocessor as the core,it has the characteristics of intelligence,good stability and high reliability.
  • Ultrathin alumina capacitive sensor is used as measuring unit with high accuracy and long calibration period.
  •  Menu lock switch:prevent misoperation and change of analyzer parameters from affecting its normal operation;
  • Alarm function:When the dew point deviates from the alarm value,the alarm will be issued.
  • Communication function:RS485 (standard) and RS232 (optional) two-way communication function can be used to communicate directly with computer or other digital communication equipment.

 

Product Parameters
Measure range: Basic type:-60~+20°C
Standard type:-80~+20°C
Extended type:-100~+20°C
Accuracy: ±3°C(-80~-60°C)
±2°C(-60~+20°C)
Repeatability: ±0.5°C
Zero drift: ±1%FS/7d
Range drift: ±1%FS/7d
Response time: At the temperature of +20°C and pressure of 1 bar, the range reaches 63% [90%] -60-*-20°C Td 5s[10s]
-20--60°C Td 45s[10min]
Power supply: 85~264VA,50/60Hz
Environment temperature: 0~45°C Environment Humidity:<90%RH
Sample temperature: 0~50°C
Sample gas pressure: atmospheric conditions+10% (Gas outlet should be atmospheric conditions)
External control signal: 1 set,control signal:9~28VDC (optional)
Analog output: Standard configuration set of 4-20mA,0-20mA and 0-10V can be set by menu
Communication:
 
RS485(standard)/RS232(optional)
 
Alarm output: 2 sets
Relay contact point capacity: 24VDC,0.2A
Inlet and outlet: 1/4 inch NPT internal thread,standard 6.5 hose joint
Instrument size: WxHxD=144mx144mmx251 mm
Installation hole size of instrument: LxW=136mmx136mm
  1. Ci-PC36 Dew Point AnalyzerCi-PC36 Dew Point Analyzer
Applications

 

  • Air separation;
  • Gas manufacturing inspection;
  • Gas humidity detection in metallurgical and smelting industries;
  • Gas humidity detection in food production process;
  • Gas humidity detection in petrochemical process;
  • Gas humidity detection in machinery manufacturing industry,etc.
FAQ

-Pressure of the measured gas:positive pressure,micro pressure or negative pressure;
-Pneumatic connection mode:

  1. 06 hose (standard);
  2. connecting by using 1/4-46 stainless steel cutting ferrule;
  3. connecting by using 1/8-<D3 stainless steel cutting ferrule;

Filter:GF2800,AF2000;
-Under the standard condition,the sensor is not suitable for natural gas operation.When ordering,the supplier should be informed whether the sample gas contains natural gas or not.



Ci-PC36 Dew Point AnalyzerCi-PC36 Dew Point Analyzer

Ci-PC36 Dew Point AnalyzerCi-PC36 Dew Point Analyzer

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2020

Suppliers with verified business licenses

Manufacturer/Factory
Registered Capital
5000000 RMB
Plant Area
101~500 square meters